Dewar Cooldown
Temperature Sensors
The temperature sensor box on the side of the dewar displays the
temperature at 7 points inside the dewar. Connect the box to the
Monitor connector on the top of the dewar to activate the sensors.
There are two types of sensors inside the dewar. Connected to the
optical bench surfaces that run at liquid (or solid) nitrogen
temperature are four diodes at points called: Saddlebags (optical
bench near bottom on LN2 can), Work Surface (near top of optical
bench), LN Shield (nitrogen shield above helium can), and GSM (grating
scan mechanism carousel).
When the switch on the temperature box is set to one of these
positions, the indicator reads the temperature in degrees Kelvin
directly. Because these diodes fail at liquid helium temperatures,
however, three resistors are attached to points called Detector
(outside of detector housing), Filter Wheel (bottom of filter wheel
cover), and 128^2 (detector substrate) which run at less than 10
K. When the switch is set to any of these sensors, refer to the
conversion chart in Figure 2.1
to convert the displayed voltage to a temperature.
Figure 2.1:
Liquid helium temperature sensor conversion chart.
Cooldown timeline
Cooling down SpectroCam-10 requires a minimum of 48 hours before the dewar is
ready for observing. The following timeline is approximate:
- 0 hours -- Vacuum Pumping.
The pumpout procedure for SpectroCam is similar to that of any cryogenic
instrument, except that the process takes longer because of the dewar's
large size.
- Attach the pump-out valve assembly, including the valve itself and the
cold trap, to the fitting on the dewar top plate, then connect a hose
from the valve to a pump (preferably the pump that normally lives
in the Cornell Lab). If you don't use the pump station, put the pinch
valve over the hose so you can regulate the pumping. Note that no cold trap is
required on the pumping station; the one mounted on the dewar valve is much
more effective because of its proximity to the dewar. We rarely use a
diffusion pump on this dewar -- a rough pump together with the valve's
cold trap is sufficient.
- Attach the vacuum gauge to its connector on the top plate to monitor
pressure. Check the dewar's pressure before opening any valves. If
the dewar is at atmospheric pressure there may be a serious leak
unless the dewar has just been closed up.
- The next step depends on the pressure inside the dewar. The dewar's
pump-out valve is of positive-pressure relief design: if the inside of
the dewar is at greater pressure than outside, the valve will open.
If the dewar is at atmospheric pressure, open the dewar valve all the
way (as it will open anyways as soon as the pump starts pulling a
vacuum). Now turn on the pump and slowly open the pump's valve over a
few minutes to gradually bring down the dewar pressure.
If the dewar is already under vacuum, turn on the pump and bring the
hoses down to a pressure below the inside of the dewar, then open the
dewar valve.
To open the dewar valve, push the shaft all the way in, turn the handle
counterclockwise until the screw stops backing out, then pull up on
the shaft. Reverse the procedure to close the valve. On the new
Palomar pump station, Valve 2 controls the rough pumping and is the
only pump station valve you need to operate.
- When the dewar pressure drops below about 2000 mTorr (typically
10 to 20 minutes after starting the rough pump), attach a liquid nitrogen dewar to
the valve's cold trap and fill it. The pressure should soon start
dropping rapidly. We usually let the dewar pump for at least a few
hours like this to evacuate the multiple layers of superinsulation and clean out
the getters as well as possible. In a hurry, the liquid nitrogen can
be filled when the dewar reaches about 200 mTorr. Pressures as low as
20 mTorr can be achieved after 3 to 5 hours, depending on the pump.
Do not let the cold trap run out of nitrogen --- its presence keeps
pump oil fumes from backflowing into the dewar.
In Case of emergency: If the pump fails, a belt breaks, or a
hose comes loose, the FIRST thing to do is close the dewar pump-out
valve. The first priority is not necessarily to ``save'' as much
vacuum as possible, but to keep pump oil and other contaminants out of
the dewar!
- 6 hours -- Liquid Nitrogen Fill
- Get a full 50 liter LN2 storage dewar ready. A complete
cooldown of SpectroCam-10 will use at least the entire 50 l.
- Close the dewar pump-out valve, making sure the shaft is pulled all
the way up once the valve has been screwed shut. Remove the small
LN2 dewar from the cold trap and allow the trap to warm up while
you proceed to the next step. You may warm the trap with a heat gun
if necessary. When the trap is warm, turn off the pump, then remove
the valve assembly from the dewar and replace the screw-on cap.
- Fill the outer can with liquid nitrogen. Use the long fill tube
that extends well into the outer can neck to minimize
cooling of the O-rings. The outer can capacity is 15 liters, and
15 -- 30 minutes are required to fill it the first time, depending on
the pressure in the storage dewar.
- When the outer can is full, put extension elbows on both
necks to vent the cold gas over the side and keep the dewar
top plate and O-rings from getting too cold.
- Attach the bayonet fitting to the inner can refill neck, and blow
dry nitrogen gas (preferably from a liquid nitrogen storage dewar)
into the refill neck for about 1 minute to clear out any water in the
tube.
- Fill the inner can with liquid nitrogen through its main fill neck.
You should detect a plume coming out of the refill line; if there is none
then the tube may be plugged. This is serious problem and must be dealt
with by warming up the inner can and clearing the plug.
- Connect the main and refill necks to a y-fitting through two
rubber hoses, and connect a third hose to the y-fitting to drain
cold gas over the side of the dewar. These connections are needed
to avoid forming condensation or a plug in the necks.
- The dewar should now cold-soak for about 24 hours. Use the
temperature sensor box to monitor the cooldown rate. The first fill
will last for 2 to 2.5 hours before the cans need to be topped off.
The hold time of the outer can once everything is cold is a few days.
About 50 liters of LN2 are required to get from fully warm to fully
cold with both cans full.
- 30 hours -- Liquid Helium fill.
- Remove the hoses and elbows from the outer can necks and the
main inner can neck. Take the hose of the inner can refill elbow
but leave the elbow in place.
- Turn the dewar upside-down in the cart to dump the cryogens in both
the inner and outer cans in preparation for the LHe fill. It is very
important to completely empty the inner can. Slowly swing the dewar back
and forth about 30 degrees and watch for spurts of LN2 coming from the
inner (Helium) can. Keep doing this until there are no more spurts (about
5 minutes).
- Using the innertube and clip assembly on a liquid helium storage
dewar, blow dry helium gas into the refill neck at a few psi for a few
minutes to clear out any liquid trapped in the thin tube.
- At least partially fill the outer can with liquid nitrogen to prevent
the dewar from warming up.
- Fill the inner can with liquid helium through the main fill neck,
using a standard transfer line. This takes about 10 minutes,
depending on the pressure in the storage dewar. We usually keep the
pressure fairly low, one or two psi. Normally the neck frosts up
pretty well almost down to its base. If the neck is iced up all the
way to its base, then a heat gun should be used to keep the O-ring
from freezing. Excessive icing may indicate a serious problem with
the dewar or transfer line. The bayonet fitting on the refill neck
should show a noticeable plume after a few minutes -- indicating that
the refill tube is not plugged.
- Attach an elbow fitting to the main neck, and connect it to the refill
line bayonet fitting with tubing and a Y-fitting. The bayonet fitting
should not be plugged because oscillations in the liquid helium will
result. Connect the third side of the Y-fitting to a long hose to
prevent backflow, and to the blowoff meter, if desired. The blowoff
meter only works when it is vertical.
- Finish refilling the outer can with liquid nitrogen if necessary.
- Put the elbows back on the outer can necks.
- The first helium fill may last only a few hours depending on
how fast the transfer was done. Following fills last for about 12
hours if the liquid nitrogen is NOT pumped, about 24 hours if pumped.
The equivalent gaseous capacity of the helium can (as measured on the
blowoff meter) is about 2400 liters.
- 32 Hours -- Initial tests:
About 2 hours after filling the helium can, the detector is cold
enough to turn on and operate (but be sure to check the temperatures
first -- the detector housing should be < 10 K [< 280 on the meter]). At this time
it may be wise to connect the detector and GSM electronics and try
some initial tests (See sections
Electronics and Cabling through
Testing Everything) to ensure
the instrument is operating properly.
- 34 Hours -- LN2 Can Pumping: If the instrument checks out OK,
then the process of pumping on the outer (liquid nitrogen) can should
be started as soon as possible. Pumping reduces the temperature of
the nitrogen from 77 to about 60 K, reducing the thermal radiation
from the interior surfaces and increasing the helium hold time from 12
to > 24 hours. The pumping process should be started at least 12
hours before mounting on the telescope to allow internal temperatures
to stabilize.
- Inspect the ``slurpee'' assembly and the pump out elbow - pinch
valve assembly, including the O-rings, pinch valve hose, and hose
clamps. Once the pumping has starting, it can be very difficult and
dangerous to break the vacuum to fix problems with these pieces, so
make sure they are OK before proceeding.
- Attach the ``slurpee'' assembly to the neck on the left side of the
dewar (the side farthest from the window). Close the valve.
- Attach the elbow fitting to the open neck on the right
side of the dewar, and connect the pinch valve hose to the pump
with a second hose.
- Close the pinch valve all the way if it's not closed already.
- Turn on the pump. Slowly open the pinch valve to begin
pumping on the nitrogen. Don't open the valve too quickly or you
might suck all the liquid out of the dewar before it freezes.
- Monitor and log the temperature sensor readings. After
several hours, the Saddlebag temperature should drop to about 60 K.
- 48 hours: By now, 42 hours after the nitrogen transfer and 16
hours after the helium transfer, the dewar should be ready to be
mounted on the telescope. Some slight additional cooling may be
noticed over the next 24 hours, but the effect of this on dewar
performance is small.