Next: Finish instrument assembly
Up: Install and verify etalon
Previous: Connect etalon to controller
To verify the operation of the etalon, we will now power up the
controller. First, check that the X, Y, and Z settings of the
dials on the left side of the controller are appropriate for the
etalon being used (see Figure 10,
Table 3). In all likelihood, unless you are using
a special etalon, the dials will not need to be adjusted.
Figure 10:
A blow-up of the etalon balance adjustment dials on the
left side of the CS-100 Fabry-Perot controller. These dials need
only be adjusted if you are using an etalon different from that
previously used with the controller. You should verify the
settings with those in Table 3.
 |
Table 3:
CS-100 Fabry-Perot controller settings for etalons commonly
used at Palomar. Note that these values may only be appropriate for
the Palomar CS-100 controller.
|
|
ET-50 |
ET-50 |
ET-70 |
|
|
#ET50-889FS |
#ET50-889FS |
#ET70-200FS |
|
|
|
(03/2001) |
|
| X |
fine |
5.10 |
5.10 |
1.55 |
| coarse |
-2 |
-2 |
-3 |
| quad |
4.06 |
4.06 |
4.18 |
| Y |
fine |
7.70 |
7.70 |
0.40 |
| coarse |
-2 |
-2 |
0 |
| quad |
3.78 |
3.78 |
3.85 |
| Z |
fine |
2.00 |
5.84 |
9.74 |
| coarse |
-4 |
-5 |
-1 |
| quad |
3.40 |
3.40 |
3.85 |
Next, make sure the operate/balance switch is on
balance. For both power on and power off, the controller
must be in balance mode. Then switch on the power to
the unit. A green light on the power switch will illuminate, as well
as a yellow ``balance'' indicator. If the meters appear normal, then
switch the operate/balance switch to operate. The yellow
balance light should switch to a green operate light, indicating that
the parallelism of the etalon plates is now being maintained by the
controller. With the offset/quadrature meter-selection switch on
offset, the three meters should read within the range ±5
volts. If not, there is a problem. With the meter switch set to
quadrature error, the reading should be close to 0 on all three
meters. Please do not turn any knobs, except the power,
operate/balance, and offset/quadrature switches, unless you are sure
that you know what you are doing. Figure 11 illustrates a properly
functioning CS-100 controller.
Note: The meters on the Palomar CS-100 are currently not
working properly in ``balance'' mode. In particular, they appear to
show a coupling between X and Z, which in fact is not
occuring. Simply set the coarse and fine settings of the dials to the
values in the table. If for some reason you think these values are
incorrect, put the etalon on a fluorescent light table and confirm
that the etalon is parallel (i.e., as you move your head side to side,
the ring radius does not increase monotonically from one side to the
other). It might in fact be a good idea to do this prior to
installing the etalon in the instrument in order to confirm proper
operation of the etalon and CS-100.
Figure 11:
A properly functioning CS-100 Fabry-Perot etalon
controller. Note that the green ``operate'' light is illuminated,
as well as the red ``disabled'' light, indicating that the
computer has control of the CS-100 interface. Note also that the
meters read approximately 0, with the meter switch set to
quadrature error.
 |
If the red out of balance light illuminates instead of the green
operate indicator, here are the recommended courses of action, in
increasing order of complexity:
- 1.
- Turn the etalon operate/balance switch back to balance,
and then back to operate.
- 2.
- Turn the etalon operate/balance switch back to balance,
and then shut off the controller power. Repeat the power up
sequence. Please always switch to the balance mode
before turning the power off or on.
- 3.
- Repeat the power cycling step, but reseat all cables while the
unit is powered off. Double check that everything is connected
properly, paying particular attention to match the small X, Y,
and Z cables to the corresponding connectors on both the
controller and the etalon.
- 4.
- Repeat the power cycling step, but check again that the X,
Y, and Z settings of the dials on the left side of the
controller are appropriate for the etalon being used (see
Table 3).
- 5.
- Contact one of the Fabry-Perot instrument support people
listed in Appendix D.
If there does appear to be a problem with the hardware, possible
solutions include calling Queensgate Instruments in New York, or
borrowing a controller from someone, such as John Trauger (JPL).
Next: Finish instrument assembly
Up: Install and verify etalon
Previous: Connect etalon to controller
Patrick Shopbell
4/23/2001