Next: Troubleshooting
Up: Verify instrument operation
Previous: Verify CCD operation
Next you should take an exposure with the calibration lamp and filter,
in order to verify the operation of the Fabry-Perot system. Due to
the faintness of the calibration lamp, it is best to turn off any
ambient lights in the dome at this time. At the 60-inch telescope,
make sure that the telescope cover has been removed, and that the
telescope has been tipped over to face a blank part of the dome. At
the 200-inch telescope, you will need to open the mirror cover at this
time. Check with the Palomar crew to make sure this is acceptable,
and to verify that the calibration lamp has been installed at prime
focus. The 200-inch telescope should be left pointing at the zenith.
First, the Fabry-Perot software must be initialized. The easiest way
to do this is via the fpsetup script:
This script initializes the serial connection between the control
computer and the etalon controller, and also defines a number of
useful logical aliases. Although the script will produce a small
number of errors related to VMS privileges, the result should be the
computer assuming control of the CS-100 etalon controller. This is
indicated by the illumination of the disabled lamp on the front
of the controller (see Figure 11). When the
controller is in this state, the front panel controls, such as the
operate/balance switch, are disabled.
To check that the computer can now control the etalon, set the etalon
spacing to the fiducial value of 100:
Any response to the fp command other than ``operating
normally'' indicates a problem with the connection between the
computer and the controller. In such event, verify the serial
connection all the way from the computer to the controller. Also make
sure that the correct serial port is being used (TTA3: at the 60-inch
telescope; TXA7: at the 200-inch telescope). John Henning may be
helpful in diagnosing problems with the serial connection.
Once the computer is controlling the etalon, place the calibration
filter in the optical path, turn on the calibration lamp, and take a
15-second exposure:
The resulting image should contain a number of illuminated rings. The
brightness of the rings depends upon the precise manner in which the
calibration lamp illuminates the dome and the choice of calibration
lamp. If the rings are too faint, you may need to try a longer
exposure. Typically, the Neon lamp requires a 15-second exposure,
while the Hydrogen lamp requires 30-60 seconds of exposure time.
The number of rings visible depends upon the free spectral range (FSR)
of the etalon in use; the Palomar ET-50 etalon will exhibit
approximately 14 rings across the frame. Assuming the etalon wedge is
being used, the rings should be centered on the edge of the frame.
Next: Troubleshooting
Up: Verify instrument operation
Previous: Verify CCD operation
Patrick Shopbell
4/23/2001