Next: Verify etalon parallelism
Up: Setup
Previous: Connect nitrogen to the
The final step in setting up the instrument is to focus the camera
lens. The camera lens must be adjusted to focus the collimated beam
coming through the etalon onto the CCD detector. This is done by
imaging the calibration lamp and adjusting the camera focus to provide
the sharpest rings possible. As you may recall from the Fabry-Perot
instrumental profile, the monochromatic rings in a Fabry-Perot image
have a width that decreases with the radius of the ring. If you are
using the Palomar ET-50 etalon, this trend is easily visible.
Consequently, the outer rings in an image are usually best for
focusing purposes.
To prepare to optimize the focus, first use the cameras command
to check that the binning pattern for the CCD is set to 1 × 1,
to optimize resolution. If you have not already done so, you should
also prepare a directory in scr:[ccd] in which to save images:
The frame command is used to initialize the frame counter; it
should probably be reset at the beginning of each night.
Next, update the on-line filter list with the names of any
non-calibration filters that have been installed:
Remember to make sure that there are no duplicate initial letters
among the filter names, and that names on the 200-inch telescope are
given in capital letters. Finally, place the calibration filter in
the optical path, turn on the calibration lamp, set the etalon spacing
to the fiducial value of 100, and start taking exposures:
Move the camera lens focus arm to one extreme of its movement. Using
the coarse scale drawn on the piece of tape fastened to the instrument
tube (Figure 17), increment the focus by a couple of
steps at a time, taking an image at each position. Use the optimal
exposure time determined from the earlier calibration images. The
RINGCHECK script or other FIGARO program may be applied to
each image to measure the width of one of the outer rings. The image
containing the minimal ring width indicates the optimal focus
position. To obtain a more accurate value, you may want to repeat
this procedure around the minimum, using smaller steps in the camera
lens focus. Always move the focus lever in the same direction when
setting the focus, to avoid possible backlash. In the case of the
Palomar ET-50 etalon and CCD #15 or #16, the optimal focus should
result in a FWHM of ~3-4 pixels for the centermost (7th) ring.
Figure 17:
The focus plate on the mounted Fabry-Perot instrument tube.
The plate may be slid horizontally along the tube, to align its
left end with markings on the coarse scale on the piece of tape.
 |
As is evident, the focus procedure is more easily accomplished with
two persons: one staying in the dome adjusting the focus of the
instrument, the other repeatedly taking an exposure at each setting.
If you are unable to find a clear minimum in the ring width, the
camera lens focus arm was probably not placed in the proper position
on the camera lens (i.e., Figure 4). Unfortunately,
the instrument must be partially disassembled to rectify this problem.
If you find this procedure necessary, also check the seating of
various components (e.g., the camera lens) when reassembling the
instrument, as this will also affect one's ability to focus the camera
lens.
Next: Verify etalon parallelism
Up: Setup
Previous: Connect nitrogen to the
Patrick Shopbell
4/23/2001