Next: Operation
Up: Setup
Previous: Verify etalon parallelism
This concludes the setup of the Fabry-Perot instrument. There are a
number of other configurations which are possible but not discussed
here. For example, the etalon wedge may be replaced with a filter
slide, which allows 2-inch square filters to be used without
vignetting. (However, due to the absence of the wedge and the
presence of the filter in the collimated beam, internal ghost
reflections will be more prevalent. Also, this method easily
incorporates only a single filter, complicating calibration
observations.) Similarly, other collimator and camera lens
combinations are possible with this system. An adaptor cable has been
purchased for incorporating etalons with ``old-style'' connectors.
Alternate configurations should be discussed with one of the contact
persons listed in Appendix D well in advance of
any observing run.
Finally, before going to dinner, you will probably want to start a
calibration data set. See Section 4.5 below for
more information on the calibration procedures.
Patrick Shopbell
4/23/2001